WIKA DI30 User Manual
Page 15
6 Commissoning, operation
WIKA operating instructions digital indicator model DI30
15
1. The cover flaps have to be opened on the right and left side. Loose the
four fixing screws, which connect the upper and lower part of the
enclosure, with a screw driver. Remove the upper part.
2. Fasten the enclosure to the wall.
3. Connecting cables have to be fed into the enclosure via the PG screws.
4. After connection the upper part of DI30 will be placed on the lower part.
Using the four fixing screws, screw the upper part to the lower part. Be
sure that the screws are completely tightened in order to attain ingress
protection IP66. Finally close the cover flaps again..
Screws for wall mounting are not part of scope of delivery.
Please use screws according to the material of the wall,
where you want to install the enclosure.