WIKA operating instructions model 54, Twin-Temp

WARNING!

Before storing the instrument (following operation),

remove any residual media. This is of particular impor-

tance if the medium is hazardous to health, e.g. caustic,

toxic, carcinogenic, radioactive, etc.

The use of liquid damping is always recommended for

temperatures near the dew point (±1 °C around 0 °C).

6. Commissioning, operation

When screwing the instrument in, the force required to do this must

not be applied through the casing, but only through the spanner flats

provided for this purpose and using a suitable tool.

6. Commissioning, operation

Installation with

spanner

When using thermowells, they must be filled with a thermal contact

medium in order to reduce the heat transfer resistance between the

outer wall of the sensor and the inner wall of the thermowell. The

working temperature of the thermal compound is -40 ... +200 °C.

WARNING!

Do not fill hot thermowells. There is a risk of the oil

spraying out!