background image

WIKA operating instructions model 232.3x.063 with 851 per ATEX

9

GB

11041781.03 03/2011 GB/D/F/E

2. Safety / 3. Specifications

Explanation of symbols

Before mounting and commissioning the pressure

gauge, ensure you read the operating instructions!

CE, Communauté Européenne

Instruments bearing this mark comply with the relevant

European directives.

ATEX European Explosion Protection Directive

(Atmosphère = AT, explosible = EX)

Instruments bearing this mark comply with the requirements

of the European Directive 94/9/EC (ATEX) on explosion

protection.

Instruments bearing this mark on the dial are safety pressure

gauges with solid baffle wall per EN 837 (S3).

3. Specifications

Pressure limitation

Steady:

3/4 x full scale value

Fluctuating: 2/3 x full scale value

Short time: Full scale value

Temperature effect

When the temperature of the measuring system deviates from the reference

temperature (+20 °C):

max. ±0.4 %/10 K of full scale value

IP Ingress protection

Enclosing case IP 54 or IP 65 as special feature (EN 60529 / IEC 529)