WIKA operating instructions model 232.3x.063 with 851 per ATEX

7

GB

11041781.03 03/2011 GB/D/F/E

Pressure gauges must be earthed via the process

connection!

WARNING!

The built-in reed switch does not have its own ignition

source.

It is a 'simple apparatus' in terms of intrinsic safety as defined

in EN 50020.

When powered by a tested and certified intrinsically safe circuit (associated

apparatus), these switches may be used in potentially explosive atmos-

pheres up to category 2 without having to take any further measures.

Operating temperatures

Ambient: -40 ... +60 °C
Medium:

The permissible medium temperature does not only depend

on the instrument design, but also on the ignition temperature

of the surrounding explosive atmosphere. Both aspects have

to be taken into account.

For permissible maximum medium temperatures see table 1.

Attention! With gaseous substances, the temperature may increase as

a result of compression warming. In these cases it may be necessary to

throttle the rate of change

of pressure or reduce the permissible medium temperature.
Table 1: Permissible medium temperature

Temperature class of the

ambient explosive atmosphere

(ignition temperature)

Maximum permissible medium

temperature (in the measuring

system)

T 6 ( > 85 °C)

+70 °C

T 5 ( > 100 °C)

+85 °C

all other temperature classes

+100 °C

2. Safety