WIKA Operating instructions pressure gauges Model 6 with 831 per ATEX

2030660 12/2004 GB/D/F

GB

Permissible vibratory stress at the mounting location

T

As a matter of principle the instruments should only be mounted at locations

without vibratory stresses

T

Where required, a decoupling from the mounting location can be achieved

e.g. by a flexible connecting line from the measuring point to the pressure

gauge and mounting via a measuring instrument bracket.

4. Alarm contacts

EC-type examination certificates

T

Standard version Model 831.XX

PTB 99 ATEX 2219 X (enclosure 2) and ZELM 03 ATEX 0128 X (enclosure 4)

Depending on the number of switches and on the case diameter either

Model SJ2-N... or Model SJ3.5-...-N... is used.

T

Safety pattern version Models 831.XX - SN or - S1N

PTB 00 ATEX 2049 X (enclosure 3) and ZELM 03 ATEX 0128 X (enclosure 4)

Depending on the number of switches and on the case diameter either

Models SJ 2-SN..., SJ 2-S1N..., SJ 3.5-SN.. or SJ 3.5-S1N... are used.

The built-in sensor type is stated on the product label of the pressure gauge.

Wiring details

T

The electrical connections should be made by qualified electricians

T

Connection of the switches via screw terminals in the terminal box

T

Conductor cross section max. 1.5 mm

2

T

The terminal assignment is stated on the connection plate at the

pressure gauge

3. Technical data ... / 4. Alarm contacts

Installation

T

Nominal position per EN 837-1 / 9.6.7 Figure 9: 90° ( )

T

Pressure connection: lower mount (LM) or lower back mount (LBM)

T

In order to avoid any additional heating, the instruments must not be

exposed to direct solar irradiation while in operation!

T

With filled versions the ventilating valve at the top of the case must be

opened prior to commissioning!