beautypg.com

Specifications, Specifications model mg-1 – WIKA MG-1 User Manual

Page 9

background image

9

WIKA operating instructions pressure transmitter, model MG-1

11609681.02 03/2013 GB/D/F/E

GB

Specifications

Model MG-1

Measuring range

bar

6

10

16

200

300

400

Overpressure safety

bar

20

20

32

500

800

800

Burst pressure

bar

25

25

160

1200

1700

1700

Measuring range

psi

100

150

200

3000

4000

5000

Overpressure safety

psi

290

290

460

7200

11000

11000

Burst pressure

psi

1450

1450

2300

17000

24000

24000

MPa, kPa and kg/cm² available; -1/0 bar or -30/0 inHg upon request

Level of cleanliness

Measuring range < 30 bar/435 psi Measuring range > 30 bar/435 psi

Breathing gas

free from oil and grease

free from oil and grease

- Residual hydrocarbons mg/m² < 1000

< 1000

Medical standard

per ISO 15001

per ISO 15001

- Residual hydrocarbons mg/m² < 550

< 220

- Particle size

µm

not applicable

on request

Industrial standard

Free from oil and grease

for oxygen per ASTM Level D/E

and DIN 19247

Free from oil and grease

for oxygen per ASTM Level D/E

and DIN 19247

- Residual hydrocarbons mg/m² < 550

< 220

High industrial standard

Free from oil and grease for

oxygen per ASTM Level C

Free from oil and grease for

oxygen per ASTM Level C

- Residual hydrocarbons mg/m² < 66

< 66

Materials

Process connection

316L and 13-8 PH

Case

316L

Electrical connection

Highly resistant glass-fibre reinforced plastic (PBT GF30)

3. Specifications

3. Specifications