Data for functional safety in the process industry

Applied standards

EN 61508:2001 and EN 61511:2004; SN29500

Device type

B

HFT

Hardware Fault Tolerance

0

SFF

Safe Failure Fraction

%

62

l

total

Total Failure Rate

FIT

101

l

no effect

Safe Undetected Failure Rate

FIT

30

l

LOW

Safe Detected Failure Rate

FIT

23

l

HIGH

Safe Detected Failure Rate

FIT

10

l

DU

Dangerous Undetected Failure Rate

FIT

38

PFD

a

Average Probability of Failure on Demand

1.66 x 10-4

T

proof

Proof Test Intervall

years

1

Suitable for Safety Integrity Level

SIL 2

(according to IEC 61511-1 section 11.4.4 under conside-

ration of proven in use)

Data for functional safety in machine building

Applied standards

ISO 13849-1:2006; SN29500

MTTF

d

Mean Time To Dangerous Failure

years

1938

Applications / Data for functional safety

GB

Applications

The pressure transmitter models IS-20-S, IS-21-S, IS-20-F and IS-21-F are suitable for
SIL2 applications in process industry for operating modes with low demand rate.
MTTF

d

values for defining the performance level (PL) are available for applications in machine

building.

Data for functional safety

Operating conditions for process industry / Operating conditions for machine building

GB

All values apply at mean ambient temperature of max. 40 °C.

The pressure transmitters are suitable for SIL2 applications assuming that the sensors
contribute approx. 35 % of the PFD

a

of the whole safety chain.

The environment of the pressure transmitter must comply with the following criteria:

„

no damage to the pressure transmitter housing

„

electromagnetic environment in accordance with EN 61326-1:2006 and EN 61326-2-3:2006

„

use of a shielded cable grounded at least at one end

The following assumptions were made for a fail-safe operation of the pressure transmitter:

„

the maximum measurement error may range from -2.5 % to 10 % of the span

„

the process connection is suitable for the measuring range

„

wetted parts are suitable for the medium

„

no pressure peaks nor static pressures are higher than the values indicated in the data
sheet

Operating conditions for the process industry

The MTTF

d

value applies at a mean ambient temperature of max. 40 °C.

The MTTF

d

value also applies to the following operating conditions:

„

no strong vibrations

„

no damage to the pressure transmitter housing

„

electromagnetic environment in accordance with EN 61326-1:2006, EN 61326-2-3:2006
and EN 61326-3-1:2008

„

the process connection is suitable for the measuring range

„

wetted parts are suitable for the medium

„

no pressure peaks nor static overpressures are higher than the values indicated in the data
sheet

Operating conditions for machine building

This manual is related to the following products: