Chemglass 30L thru 50L Jacketed User Manual
Page 5
www.cglifesciences.com
Inlet/Outlet Jacket Connections
The stainless steel adapters are assembled and attached when shipped. Use the following instructions should you need to
remove or adjust these components.
Components Needed for Inlet/Outlet:
2ea CG-1969-A
S.S. Adapters, 1”Beaded Pipe to M16x1.0 Male or M30 x 1.5 Male
2ea CG-1968-66
1”Beaded Pipe Couplings
1. Reactor jacket comes standard with 1”glass beaded pipe on the inlet/outlet.
2. Loosen the nut on the beaded pipe coupling. Wet the PTFE/Viton liner inside
the coupling and attach one side to the 1”beaded pipe on the jacket.
3. Insert the beaded pipe adapter (CG-1969-A) into the other side of the coupling.
4. Tighten the 1”beaded pipe coupling to 35in-lbs using a torque wrench.
MAX JACKET PRESSURE = 12 PSI ON 30L JACKETED REACTOR CG-1968-61
MAX JACKET PRESSURE = 12 PSI ON 50L SHORT JACKETED REACTOR CG-1968-95
MAX JACKET PRESSURE = 12 PSI ON 50L TALL JACKETED REACTOR CG-1968-81
MAX T = 60°C
TEMPERATURE RANGE = -60 to 200°C
The Chemglass High Flow Manifold
System alleviates excessive mechanical
forces on the inlet/outlet and prevents
over pressurizing with a pressure relief
valve. This manifold is available
separately, see CG-1969-M.