measurement keys will exit without changing the setting. User
mode “Linc” is now complete. The User mode will now use the
calibration gain constant used for the calibration of the
Standard Solution as outlined above.
C. Canceling User mode Calibration “Linc”
The Ultrameter II must be in “User” linked mode in order to cancel the
“Linc” (ref. SOLUTION SELECTION, pg. 12).
1.
Press “Linked” measurement key
,
or
.
Two solution icons will be shown in the left side of display -
“User” and another, e.g., “KCl”.
2.
Press
key until the menu “Linc” appears (see Figure 27).
3.
Press
key; the instrument will display both “SEL” and
the “User” Icon.
4.
Press
key until “User” is the only solution icon being
displayed.
5.
Press
key.
6.
The User mode calibration “Linc” has now been canceled.
NOTES:
1.
To maintain repeatability, use the same standard solutions for
future calibrations.
2.
Calibration of the Ultrameter II Gain Factor for User mode is
not available when the calibration linc has been established.
The other calibration functions (i.e., Temperature Compensation
%/C settings and TDS Ratio settings) are still intact. To
perform a calibration of the User mode as described in User
Calibration Conductivity/TDS, pg. 16, the User mode Linc should
be canceled. See above, “Canceling User mode calibration
“Linc””.
3.
Once a “Linc” has been established for User mode , the “Linc”
will apply to all measurement modes using User solution
selection (i.e., TDS/User, Cond/User or Res/User).