Myron L 6Pfc and 4P User Manual

Page 18

background image

14

2. Disabling Temperature Compensation

a.

Select User mode (ref. Procedure to Select a Solution, pg. 13).

b.

With “User” selected, press

. If the display does not

show .00%/°C, hold long enough to bring the tempco to

.00%/°C (see Figure 3).

Figure 3

°C

% /

User

COND

c.

Press

twice

(3 times if in TDS mode).

Temperature compensation

is now disabled (=0) for

measurements in User mode.

3. User Programmable Conductivity to TDS Ratio

This feature allows you to select a custom conductivity to TDS conversion

ratio within the range of 0.20-7.99 for User mode measurements.
To determine the conversion ratio for a custom solution of known

TDS ppm value, measure the solution conductivity at 25ºC with the

Ultrameter II and divide the ppm value by the µS value. For example,

a solution of known 75 ppm TDS and measured 100 µS conductivity at

25ºC would have a conversion ratio of 75/100 or 0.75. Enter the new

conversion ratio as follows:
a.

While in User mode, press

.

b.

Press

twice (to skip

Figure 4

RATIO

User

TDS

over tempco adjustment) and

RATIO” will appear (see

Figure 4).

c.

Adjust with

or

key until new conversion

ratio is displayed.

d.

Press

twice (to skip over calibration adjustment) to

accept new conversion ratio. You are now ready to measure

samples with the new conductivity/TDS ratio.

In these first six sections, you have learned

all you need to make accurate measurements.

The following sections contain calibration,

advanced operations and technical information.