Myron L 6Pfc and 4P User Manual

Page 32

background image

28

B. Setting User mode Calibration “Linc”

The Linc function sets or “links” the calibration gain factor of a Standard

Solution to the User solution mode. Once set, the “Linc” will stay intact

with future calibrations unless the Linc has been canceled. For more

information on canceling the User mode Calibration Linc refer to the

section “Canceling User mode Calibration “Linc””, pg. 29.

Follow the steps below to set either the KCl, NaCl or 442 calibration

factor to the User solution mode.
1.

Press measurement key desired to be “Linked”, i.e.,

,

or

.

2.

Place the Ultrameter II in

Figure 27

User mode (ref. SOLUTION

SELECTION, pg. 12, for

selecting the User mode).

3.

Press

arrow key until

the menu “Linc” appears

(see Figure 27).

Figure 28

User

4.

Press

key. The

instrument will display “SEL

and the “User” Icon (see

Figure 28).

Any additional display of KCl,

NaCl or 442 icons indicates a “Linc”

Figure 29

KCl

User

between the User solution and the

other solution displayed.

5.

Press

or

keys

to select a Standard

Solution to be linked to the

User mode calibration

constant (see Figure 29).

User” is linked to “KCl”.

If none of the Solution Selection icons are displayed (i.e., KCl, NaCl or

442), nothing has been linked to User mode.

6.

Press

key to accept the setting. Pressing any of the