Myron L PS6FCE User Manual
Page 15

10
11
Figure 1
3.
Press the
or
keys to toggle between mV (standard
ORP mode) and FC
E
ppm. The setting chosen is displayed. (See
Figure 2.)
Figure 2
PPM
4.
Press any parameter key to exit ORP unit preference selection
or let the unit time out. ORP unit preference will be saved.
2. Measuring ORP
1.
Ensure the PS6FC
E
is in ORP mode (ref. ORP/FC
E
Mode
Selection, pg. 10).
2.
Remove protective cap by rotating while grasping and pulling up.
3.
Rinse sensor well and cell cup 3 times with sample to be
measured. Shake out each sample to remove any residual liquid.
4.
Refill both sensor well and cell cup with sample.
5.
Press
.
6.
Take reading.
7.
Press MS to store reading in memory, if desired.
IMPORTANT: After use, fill pH/ORP sensor well with Myron L