beautypg.com

Myron L 762 User Manual

Page 14

background image

Component Identificaiton, Calibration and Preventive Maintenance

STEP 1 Set the controls of a test meter to read at least 20 mA

and connect it between the positive terminals of the
Transmitter and Receiver as shown in Fig. 4-7 (4-20
mA option) and Fig. 4-8 (4-20D option).

NOTE:

Receiver connected to 420D option should have an
input impedance less than 500 ohms.

STEP 2 Press and hold the “SET POINT CHECK” switch. It

must be depressed during all the following adjustment
steps.

STEP 3 Turn the Set Point trimmer (R31 ) (see Fig. 4-1 ) until

the meter displays a reading of 0.00.

STEP 4 Turn the 4-20 mA Null trimmer (R23) (see Fig. 4-6) until

the test meter indicates 4 mA.

STEP 5 Readjust the Set Point trimmer (R31 ) until the meter

displays 1999 (decimal point omitted).

STEP 6Turn the 4-20 mA Full Scale trimmer (R20) (see Fig. 4-

6) until the test meter indicates 20 mA.

STEP 7 Repeat STEP 3 through STEP 6, readjusting as

required.

STEP 8 Reset the set point to the desired setting. Release the

“SET POINT CHECK” switch.

4.4

PREVENTIVE CARE

Although designed to minimize operating problems, the Myron L
Company recommends that the following Preventive Care
procedures be observed.

1

Try to prevent exposure to excessive heat and
moisture.

2

The Monitor’s main AC power source must be
protected against excessive voltage spikes.

3

Take care not to damage Monitor during handling.

NOTE:

Daily, weekly or monthly maintenance schedules are based
upon the frequency of use and the severity of the Monitor’s
environment and operating conditions.

STEP 1 Repeat the Monitor’s Check-Out procedures and/or

isolate possible troubleshooting symptoms.

STEP 2 Check all cable connections to ensure that they are

free of moisture and contamination.

STEP 3 Inspect and replace damaged component boards and

cable assemblies.

STEP 4 Remove and clean sensor with soap and water. Rinse

thoroughly.

12

This manual is related to the following products: