beautypg.com

Nor-Lake Lighted Chambers User Manual

Page 5

background image

- 5 -

HUMIDITY INSTALLATION (if applicable)

The water supply should have a dedicated shutoff valve and regulator installed. Connect

the water supply to the humidifier (the operating pressure range is 10-60 psi). For best

results, provide a source of reverse osmosis water with a resistance of .05 to 1.0mΩ, which

will reduce the scale build up and cleaning frequency. Unit is equipped for water supply with

1/8” Male Pipe Threads. The humidifier reservoir water level may be adjusted by using the

adjusting thumb screw located on the float valve (see below).

The humidifier is a precision built instrument

that, given proper care, will provide years of

dependable service. Cleaning the humidifier

periodically is all that is required to insure

proper operation. Depending on local water

conditions, mineral solids and other matter

may accumulate in various parts of the unit.

These accumulations must be removed so

that water flows freely for efficient operation. Failure to clean the humidifier may result in

malfunction which will lead to repairs.

If higher humidity levels are unobtainable, loosen the

reducing coupling clamp and remove reducing coupling

boot and reducer PVC.




DRAIN INSTALLATION

The Chamber has a floor drain in the bottom of the

cabinet. A silicone drain stopper has been placed in this

drain. The drain stopper will prevent undesirable liquids from running into the drain. The

stopper can be removed if any possible liquids flowing into this drain are acceptable to go

to the floor drain.

An 18” length of 1/2" I.D. drain hose is shipped inside cabinet. Attach hose to drain stub

under cabinet with supplied clamp and run to an atmospheric floor drain. May substitute

longer drain hose as needed. Humidifier units will require a drain for overflow.