Nor-Lake Lighted Chambers User Manual
Page 4
- 4 -
Shelving Installation
– Shelving and Lighting comes installed for your desired light option. To
change shelf locations disengage lock as shown in Detail A, slide bracket up and remove. To install
insert shelf bracket tabs into pilaster and push down. Slide lock fully into hole.
Remote Alarms Contacts Access
- The factory installed Remote alarm
contacts access box is located at the top of the cabinet behind the
microprocessor control.
1. Remove the cover to access the terminal connections.
2. Select and knock-out a hole to run field leads into electrical box
terminals.
3. The terminal block in the electrical box is
labeled for “normally open” and “normally
closed” activation. End user is
responsible for proper field installation.
Terminal connections are rated for class
II circuits only per NEC table 11(A).
(Limited power source less than 30vac 8
Amp. max, see applicable notes in NEC).
2-10 volt DC Output
, is located at the top of the
control box behind the microprocessor control,
connect wires as per label (4-20 ma optional).
RS485 port
- (Optional) terminal board for
RS485 port is located behind the cabinet façade, on the left hand side of the control box,
connect wires as per label.
Duplex/ or European Outlet
- (Optional) is located near the top of the interior cabinet
back. This is a 15 amp 115 volt duplex PN 88010300 with its own power supply cord.
Duplex outlet power cord is wrapped up on the back of the cabinet top. Outlet is not inter-
wired thru main cabinet supply and must be plugged into a 115 volt/ or 220/240V power
supply. Locate cabinet within 8 feet of the wall receptacle.
Access Port
- (Optional) is provided with a spring loaded cover on the right hand side of
the cabinet.