Standard Commands for Programmable Instruments (SCPI)
95
*TRG
Meaning and Type
Trigger
Device Trigger
Description
This command, which is essentially the same as the Group Execute Trigger (
TRIG[:IMM]
signals, generates a trigger to the power supply.
Command Syntax
*TRG
*TST?
Meaning and Type
Test
Device Test
Description
This query causes the power supply to do a self test and report any errors.
Query Syntax
*TST?
Returned Paramters
0
Indicates power supply passed self-test.
Nonzero
Indicates a self-test failure.
*WAI
Meaning and Type
Wait to Continue
Device Status
Description
This command instructs the power supply not to process any further commands until all pending operations are completed.
"Pending operations" are as defined under the
*OPC
command.
*WAI
can be aborted only by sending the power supply a
GP-IB
DCL
(Device Clear) command.
Command Syntax
*WAI
ABOR
This command cancels any pending triggered levels stored by the
CURR:TRIG
or
VOLT:TRIG
commands. The pending
level is set equal to the corresponding immediate value.
ABORt
also resets the WTG bit in the Operation Condition status
register (see "Status Reporting"). If
INIT:CONT ON
has been programmed, the trigger system rearms itself immediately
after
ABORt
, thereby setting WTG. ABORt is executed at power turn on.
Command Syntax
ABORt
Examples ABOR
ABORT