Syst:err, Syst:lang – Agilent Technologies 6028A User Manual
Page 102
![background image](/manuals/45568/102/background.png)
Standard Commands for Programmable Instruments (SCPI)
102
Command Syntax
STAT
us
:QUES
tionable
:NTR
ansition
STAT
us
:QUES
tionable
:PTR
ansition
Parameters
0 to 32727
Default Value
0
Examples
STAT:QUES:NTR 16 STATUS:QUESTIONABLE:PTR 512
Query Syntax
STAT:QUES:NTR? STAT:QUES:PTR?
Returned Parameters
(Register value)
SYST:ERR?
Returns the next error number followed by its corresponding error message string from the remote programming error
queue. The queue is a FIFO (first-in, first-out) buffer that stores errors as they occur. As it is read, each error is removed
from the queue. When all errors have been read, the query returns 0,N0 ERROR. If more errors are accumulated than the
queue can hold, the last error in the queue will be -350,T00 MANY ERRORS.
Query Syntax
SYST
em
:ERR
or
?
Returned Parameters
Examples
SYST: ERR? SYSTEM: ERROR?
SYST:LANG
Switches the interface between its SCPI (TMSL) command language and it compatibility language. Note that the earlier
parameter "TMSL" must be used in place of "SCPI". Sending the command causes:
♦
The alternate language to become active and to be stored in nonvolatile memory.
♦
The power supply to reset and clear the storage and error registers.
NOTE
After sending the SYST:LANG command, do not send another command or turn the supply off for at least
five seconds
.
If the power supply is shut off, it will resume operation in the last-selected language when power is restored.
Command Syntax
SYST
em
:LANG
uage
{TMSL|COMP
atibility
}
Syntax is the same, regardless of the present language
Parameters
or
COMPatibility
Default Value
TMSL or last selected language
Examples
SYST:LANG TMSL SYSTEM:LANGUAGE COMPATIBILITY
Query Syntax
SYST
em
:LANG
uage
?
Returned Paramters
or
COMP