Command Syntax
*ESE
Parameters
0 to 255
Default Value
(See
*PSC
)
*RST Value
0
Example
*ESE 129
Query Syntax
*ESE?
Returned Parameters
Related Commands
*ESR? *PSC *STB?
*ESR?
Meaning and Type
Event Status Register
Device Status
Description
This query reads the Standard Event Status Event register. Reading the register clears it. The bit configuration of this
register is the same as the Standard Event Status Enable register (
*ESE
). See "TMSL - Status Reporting" for a detailed
explanation of this register.
Query Syntax
*ESR?
Returned Parameters
(Register binary value)
*IDN?
Identification
Query
Meaning and Type
Identification
System Interface
Description
This query requests the power supply to identify itself.
Query Syntax
*IDN?
Returned Parameters
String
Information
Agilent Technologies
Manufacturer
xxxxA
4-digit model number followed by a letter suffix
nnnnA-nnnnn
10-character serial number or
0
fx.xx.sx.xx,px.xx
Revision levels of firmware
Example AGILENT
TECHNOLOGIES,6632,2701A-00101,f5.06,s5.02,p5.06
*OPC
Meaning and Type
Operation Complete
Device Status