Agilent Technologies 6028A User Manual
Page 76
100 VAC Input Power Option 100
76
Section ll Manual Changes
For Agilent Models 6030A, 6031A, 6032A and 6035A: on page 2-7 (AC line Impedance Check, step b), where the
maximum output voltages are tabulated, change the voltages as shown below:
6030A change 65 V to 50 V
6031A change 8 V to 6 V
6032A change 22 V to 13.5 V
6035A change 220 V to 150 V
Section lll Manual Changes
On page 3-10 (Overvoltage Protection), change 107% to 90%.
On page 3-11 (Monitor Signals), change “0-5 Volts” to ‘’0-4.5 Volts’ ‘ .
In Table 3-6, change the soft voltage limits as follows:
Model
from:
to:
6030A
204.75
174.03
6031A
20.475
17.40
6032A
61.425
52.21
6035A
511.875
435.10
6033A
20.475
17.40
6038A
61.425
52.21
In table 3-7, GP-IB Commands, change the boxes in the second column (*Range **Response) with voltages tabulated in
tables a, b, c, or d as shown below.
Change the column for VSETx, VSETxV, VMAXx, and VMAXxV to set a.
Change the column for VSETxmV, and VMAXxmV to set b.
Change the column for ISETxA, ISETxA, IMAXx, and IMAXxA to set c.
Change the column for ISETxmA, and IMAXxmA to set d.
Agilent Model
Set a
Set b
Set c
Set d
6030A
0-175.00 V
0-175000 mV
0-17.403 A
0-17403 mA
6031A
020.45 V
0-20425 mV
0-122.85 A
0-122850 mA
6032A
0-51.495 V
0-51495 mV
0-51.1875 A
0-51187.5 mA
6033A
0-17.500 V
0-17500 mV
0-30.7125 A
0-30712.5 mA
6035A
0-437.50 V
0-437500 mV
0-5.119 A
0-5119 mA
6038A
0-51.495 V
0-51495 mV
0-10.2375 A
0-10237.5 mA