P2546A-L Anemometer

Nonlinearity:

< 0.04 m/s

Temperature Influence
(-15° to 60°C):

< 0.05 m/s

6.1 Calibration

Standard:

U=A0+B0×f ,

Where:

U=Wind speed in m/s

f = Output frequency in Hz

A0 = 0.27 m/s

B0

= 0.620 m

6.2 Switching Characteristics

Signal Type:

potential free contact closure

Duty Cycle:

40% to 60%

Maximum Switching Voltage: 30 V

Maximum Recommended
Switching Current:

10 mA

Series Resistance:

330 Ω, 1 W

Operating Temperature:

–35° to 60°C

7. Installation

Locate wind sensors away from obstructions (e.g., trees and building). As a
general rule, there should be a horizontal distance of at least ten times the
height of the obstruction between the wind set and the obstruction. If mounting
the sensors on the roof of a building, the height of the sensors above the roof
should be at least 1.5 times the height of the building. See Section 10

for a list

of references that discuss siting wind speed and direction sensors. For power
performance applications, refer to IEC 61400-12-1 which specifies the
mounting and location of anemometers.

7.1 Wiring

Connections to Campbell Scientific dataloggers are given in TABLE 7-1 and
TABLE 7-2.
When Short Cut program generator software is used to create the
datalogger program, wire the sensor to the datalogger as directed by the wiring
diagram created by Short Cut.

7