beautypg.com

2 hydrosense modes, Specifications – Campbell Scientific HydroSense® Soil Water Content System (CS620, CD620) User Manual

Page 6

background image

HydroSense Soil Water Measurement System (CD620, CS620)

1.2 HydroSense Modes

The Water Content Measurement Mode displays the measurement result as
percent volumetric water content and shows the period of the probe output in
milliseconds.

In the Water Deficit Mode, HydroSense measurements are taken at lower and
upper water contents chosen by the user and stored in memory as reference
values. The reference values are then applied to subsequent measurements to
determine the amount of water that must be added to bring the soil to the upper
water content. The relative water content based on the reference values is also
displayed.

See Sections 4 and 5 for detailed operating information.

2. Specifications

CD620 HydroSense Display Unit

Measurement

% Volumetric Water Content

Resolution 1.0%

Environmental Splash

resistant

Display

16 character, 2 line LCD

Keypad

Two button membrane keypad

Power

2 x AAA 1.5 VDC alkaline batteries

Current Consumption

Sleep

<50

μA

Idle

1.5

mA

Measurement

70

mA

Battery Life

Approximately 12 months

Dimensions

120 (w) x 73 (d) x 24 (h) mm (4.7” x 2.9” x 0.9”)

Weight

166 g (7 oz) includes batteries

CS620 Water Content Reflectometer

Accuracy

±3.0% volumetric water content with electrical
conductivity <2 dS m

-1

.

Range

0% to saturation*

Stabilization time

Instantaneous

Response time

<50 milliseconds

Power requirement

100 milliamps at 5

±0.03 volts

Output

Square wave with ±2.5 VDC amplitude

Dimensions

Head

105 x 70 x 18 mm (4.1” x 2.8” x 0.7”)

Rods

120 or 200 mm length , 5 mm diameter, 32

mm spacing