beautypg.com

Maintenance, 1 cleaning, 1 when to clean – Campbell Scientific CS526 ISFET pH Probe User Manual

Page 18: Cleaning, When to clean

background image

CS526 ISFET pH Probe

Problem:

Output signal is at its maximum value.

Possible reasons:

o Probe is not in fluid.

o Chip is polluted.

o Diaphragm is polluted.

o Chip is scratched.

Suggestions:

o Put probe in fluid.

o Clean probe (Section 9.1, Cleaning).

o Probe cannot be fixed if chip is scratched.

Problem:

Probe response is very slow.

Possible reason:

o Diaphragm is chipped or polluted.

Suggestion:

o Clean probe (Section 9.1, Cleaning).

Problem:

Probe signal is drifting.

Possible reasons:

o Diaphragm can be dried out.

o Chip is scratched.

Suggestions:

o Soak probe for 10 minutes in saturated KCl solution (pn 16349).

o Probe cannot be fixed if chip is scratched.

9. Maintenance

The CS526 needs to be periodically cleaned and calibrated

to ensure accurate readings and proper operation.

The CS526 has no user-serviceable parts.

9.1 Cleaning

Proper maintenance of a probe is important. If the probe is not properly and

regularly cleaned, the probe can malfunction due to a contaminated diaphragm

or ISFET chip. Contamination on/or blockage of the sensor and reference

electrode diaphragm surface is the most likely cause for probe failure.

Probe wear is another cause for probe failure. Probe wear is often, but not

necessarily, preceded by a period of declining calibration slope values. Probe

wear is dependent on how the probe is used and stored (see Section 9.2,

Storage). Worn probes need to be replaced.

9.1.1 When to Clean

Often the probe should be cleaned daily, but the appropriate cleaning frequency

is dependent on the type of sample being measured.

CAUTION

10