CS450/CS455 Pressure Transducer

5.2 RS-232

TABLE 5-2. RS-232 Wiring


Color

CS450/CS455
Function


Connection


RS-232 9-pin

Red +12vdc

Power

Source

Black

Power Ground

Power Ground

White

RS-232 Tx (Output)

Transmit

Pin 2 Rx (Input)

Blue

RS-232 Rx (Input)

Receive

Pin 3 Tx (Output)

Yellow

Digital Ground

Ground

Pin 5 GND

Clear Shield

GND

Ground

5.3 A150/CWS900

The A150 desiccated case allows the CS450/CS455 to be connected to the
CWS900 wireless sensor and then used in a wireless network. FIGURE 5-1
shows the CS450 wired to the A150’s terminal block. The 26972 cable is
included with the A150.

BLUE

RED

CLEAR

GREEN

1

2

BLACK

3

4

WHITE

5

6

BLACK

#26972 Cable

#19520
CWS900 Connector

WHITE

BLUE

BLACK

YELLOW

CLEAR

RED

TUBE

N.C.

N.C.

Terminates in A150 Enclosure

A150

Pressure Sensor
CS450

Sensor TX (Out)

Sensor RX (In)

Power GND

Digital GND

Shield

Power

Vent

(Pin #)

FIGURE 5-1. A150 wired for CS450 pressure sensor and CWS900

connector

6. Programming

6.1 Using SCWIN

SCWIN is the easiest and typically the preferred method for programming the
datalogger. SCWIN generates a wiring diagram that shows how to connect the
CS450/CS455 to your Campbell Scientific datalogger.

The sections that immediately follow are for CRBasic and Edlog.

NOTE

16