CS420-L and CS425-L Pressure Transducers
6: End (P95)
; Deactivate Multiplexer
7: Do (P86)
1: 51
Set Port 1 Low
; Adjust sensor readings for multiplier and
; offset:
8: Scaling Array (A*Loc + B) (P53)
1: 1
Start
Loc
[DEPTH_FT]
2: 2.2975 A1
;Multiplier
#1
3: .36781 B1
;Offset
#1
4: 1.0560 A2
;Multiplier
#2
5: .4011
B2
;Offset
#2
6: 3.0001 A3
;Multiplier
#3
7: .2103
B3
;Offset
#3
8: 1.9821 A4
;Multiplier
#4
9: 1.0029 B4
;Offset
#4
; Record average depth every hour.
9: If Time is (P92)
1: 0000
Minutes (seconds --) into a
2: 60
Minute Interval (same units as above)
3: 10
Set output flag high (Flag 0)
10: Real Time (P77)
1: 110
Day,
Hour/Minute
11: Average (P71)
1: 4
Reps
2: 1
Loc
[DEPTH_FT]
9. Maintenance
Periodic evaluation of the desiccant is vital for keeping the vent tube dry. To
assess the effectiveness of the desiccant, use one of the following:
• An indicating desiccant that changes color when it’s losing its drying
power
• An enclosure humidity indicator such as our #6571 humidity indicator
card
9.1 Every Visit, At Least Monthly
• Collect data
• Visually inspect wiring and physical conditions