beautypg.com

2 instruction 9 (6-wire full bridge w/excitation), 1 wiring, 2 edlog program – Campbell Scientific CS420 and CS425 PDCR 1830-8388 Pressure Transducer User Manual

Page 18

background image

CS420-L and CS425-L Pressure Transducers

7.2.2 Instruction 9 (6-Wire Full Bridge w/Excitation)

7.2.2.1 Wiring

Datalogger

CS420-L/CS425-L

Excitation

E1

Red (Supply +)

Input

H1

Orange (Monitor +)

Input

L1

Black (Monitor -)

Input

H2

Yellow (Output +)

Input

L2

Blue (Output -)

G

White (Supply -)

G

Clear

(Shield)

7.2.2.2 Edlog Program

; Measure one 6-wire DRUCK Sensor.

* Table 1 Program
01: 5

Execution

Interval

(seconds)


1: Full Bridge w/mV Excit (P9)
1: 1

Reps

2 25*

5000 mV, 60Hz Rejection, slow Ex Range

3: 22**

50 mV, 60 Hz Rejection, Slow Br Range

4: 1

DIFF

Channel

5: 1

Excite all reps w/Exchan 1

6: 5000

mV

Excitation

7: 1

Loc

[DEPTH_FT]

8:

2.3529 Mult

9: -0.49

Offset

*For 21X use Range Code 5
**For 21X use Range Code 3

; Every 60 minutes output the array ID, time,
; and average water depth in feet

2: If time is (P92)
1: 0000

Minutes

into

a

2: 60

Minute

Interval

3: 10

Set output flag high


3: Real Time (P77)
1: 110

Day, Hour/Minute (midnight = 0000)


4: Average (P71)
1: 1

Reps

2 1

Loc

[DEPTH_FT]

14