2 accuracy, 3 general, Accuracy – Campbell Scientific CS106 Barometer User Manual
Page 13: General

CS106 Barometric Pressure Sensor
6.2 Accuracy
Total Accuracy***:
±
0.3 mb @ +20 °C
±
0.6 mb @ 0 to +40 °C
±
1 mb @ –20 to +45 °C
±
1.5 mb @ –40 to +60 °C
Linearity*:
±
0.25 mb @ 20 °C
Hysteresis*:
±
0.03 mb @ 20 °C
Repeatability*:
±
0.03 mb @ 20 °C
Calibration Uncertainty**: ±0.15 mb @ 20 °C
Long-Term Stability:
±
0.1 mb per year
* Defined as ±2 standard deviation limits of end-point non-linearity,
hysteresis error, or repeatability error
** Defined as ±2 standard deviation limits of inaccuracy of the working
standard at 1000 mb in comparison to international standards (NIST)
*** Defined as the root sum of the squares (RSS) of end-point non-linearity,
hysteresis error, repeatability error and calibration uncertainty at room
temperature
6.3 General
Dimensions:
9.7 x 6.8 x 2.8 cm (3.8 x 2.7 x 1.1 in)
Weight:
90 g (3.2 oz)
Housing Material:
ABS/PC blend
Supply Voltage:
10 to 30 Vdc
Supply Voltage Control:
When the internal jumper is closed, the CS106
is on continually. When the jumper is open,
the CS106 can be turned on/off with 5 Vdc/
0 Vdc.
Supply Voltage Sensitivity: negligible
Current Consumption:
<4 mA (active); <1 µA (quiescent)
Output Voltage:
0 to 2.5 Vdc
Warm Up Time:
1 s
Pressure Fitting:
barbed fitting for 1/8 in I.D. tubing
Overpressure Limit:
2000 mb
5