AM16/32A Relay Analog Multiplexer

Multiplexer to Datalogger wiring - If the reference junction is at the
datalogger, then the wire that connects the COM ODD H, COM ODD
L, and COM EVEN H terminals to the datalogger should be of the
same composition as the high side of the thermocouples. Also, the
wire that connects COM EVEN L to datalogger ground should be of
the same composition as the low side of the thermocouples.

If the reference junction is at the AM16/32A (CSI 107 thermistor,
RTD, etc.), then copper wire should be used to connect COM
terminals to the datalogger.

6.6.3 Differential Thermocouple Measurement

AM16/32A panel switch set to “2x32” mode.

Multiple

handled in one of two ways. If a reference junction (107 thermistor,

copper wire may be run

xer and the datalogger

6.7 Mixed Sens

In applications where sensor types are mixed, multiple hook-up

possible. Please

consult CSI for application assistance if you need to multiplex

6.7.1 Mixed Sensor Example: Soil Moisture Blocks and

Thermocouples

AM16/32A panel switch set to “4x16” mode.

In this example, 16 thermocouples and 16 soil moisture blocks will be
multiplexed through the AM16/32A. One thermocouple and one soil
moisture block are input into each SET.

xer to Datalogger wiring - The wire types here can be

or RTD, etc.) is at the AM16/32A, then one
between the COM terminals of the multiple
input channel.

If the reference junction is at the datalogger, then matching
thermocouple wire should be run between the COM terminals of the
multiplexer and the differential input channel on the datalogger
(observe TC wire polarity).

or Types

configurations and programming sequences are

markedly different sensor types in your application.