AM16/32A Relay Analog Multiplexer

Low level single-ended measurements are not
recommended in 21X applications where the 21X's
internal 12VDC supply is used to power the multiplexer
or other peripherals (see Section 4.1.4).

5000

"2 X 32" Mode

COM ODD H

COM ODD L

ODD H

ODD L

(+) SENSOR

(-)

COM

AM16/32

MUXSIGNAL

SHIELD

SENSOR SHIELD

21X/

CR7


CR10(X)

CR23X/CR3000/

CR800/CR850/

CR1000/CR5000

H H

H

AG

G

FIGURE 10. Single-ended Measurement without Excitation

5000

"4 X 16" Mode

COM ODD H

COM ODD L

ODD H

ODD L

(+) SENSOR

(-)

COM

AM16/32

MUXSIGNAL

SHIELD

SENSOR SHIELD

21X/

CR7


CR10(X)

CR23X/CR3000/

CR800/CR850/

CR1000/CR5000

H H

H

L L

L

G

FIGURE 11. Differential Measurement without Excitation

6.2 Differential Analog Measurement without Sensor

Excitation

at

th panel

tch set to “4x16” mode. Sensor shields are connected to the input

Sensor to Multiplexer wiring - Up to two differential sensors th
don't require excitation may be connected to one input SET wi
swi

” terminals.

Multiplexer to Datalogger wiring - The two pairs of COM terminals
(O D H, ODD L

D

and EVEN H, EVEN L) are connected to two pairs

of differential analog inputs at the datalogger. Observe H to H and L

ential sensors in SETs of two with appropriate

programming.

NOTE

to L from sensor to multiplexer to analog input. In “4x16” mode up to
32 differential sensors can be measured by two differential datalogger
channels in this way.

With panel switch set to “2x32” mode, one differential input can
measure up to 32 differ

23