4WPB100, 4WPB1K PRT Bridge Terminal Input Modules

9

With the PRT at 0

°C, R

s

=R

o

. Thus, the above result becomes R

o

/R

f

, the

reciprocal of the multiplier required to calculate temperature, R

f

/R

0

. By

making a measurement with the PRT in an ice bath, errors in both R

s

and R

o

.

can be accounted for.

To perform the calibration, connect the PRT to the datalogger and program the
datalogger to measure the PRT with the 4 wire half bridge as shown in the
example section (multiplier = 1). Place the PRT in an ice bath (@ 0

°C;

R

s

=R

0

). Read the result of the bridge measurement. The reading is R

s

/R

f

,

which is equal to R

o

/R

f

since R

s

=R

o

. The correct value of the multiplier, R

f

/R

0

,

is the reciprocal of this reading. For example, if the initial reading is 0.9890,
the correct multiplier is: R

f

/R

0

= 1/0.9890 = 1.0111.