beautypg.com

3 cr7, 4 cr9000x – Campbell Scientific 4WPB100, 4WPB1K PRT Terminal Input Modules User Manual

Page 10

background image

4WPB100, 4WPB1K PRT Bridge Terminal Input Modules

6

4.3 CR7

1: Full Bridge w/mv Excit (P9)
1: 1

Reps

2: 3

± 15 mV Slow Range

3: 3

± 15 mV Slow Range

4: 1

In

Card

5: 1

DIFF

Channel

6: 1

Ex

Card

7: 1

Ex

Channel

8: 1

Meas/Ex

9: 4070

mV

Excitation

10: 1

Loc [ Rs_Ro ]

11: 1.0

Mult

12: 0.0

Offset


2: Temperature RTD (P16)
1: 1

Reps

2: 1

R/RO Loc [ Rs_Ro ]

3: 2

Loc [ Temp_C ]

4: 1.0

Mult

5: 0.0

Offset

4.4 CR9000X

'CR9000X Datalogger
Public Rs_Ro, Temp_F

DataTable (Temp_F,1,-1)
DataInterval

(0,0,0,10)

Sample

(1,Temp_F,FP2)

EndTable

BeginProg
Scan (1,mSec,0,0)
BrHalf4W (Rs_Ro,1,mV50,mV50,4,1,5,7,1,4070,True,True,30,40,1.0,0)
PRT

(Temp_F,1,Rs_Ro,1.8,32)

CallTable

Temp_F

NextScan
EndProg