1.1 Specifications

Temperature Range: -40° to +135°C

Survival Range: -50° to +140°C

110PV Temperature Uncertainty

-40° to 70°C: ±0.2

°C

71° to 105°C: ±0.5

°C

106°

to

135°C:

±1

°C

Time Constant (average):

Test

τ

Still Air 252 seconds

Surface

25

seconds

Water Submersion Depth: 50 ft (21 psi)

Linearization Error: Steinhart & Hart equation; maximum error is 0.0024°C at
-40°C.

Maximum Cable Length: 1000 ft

Disk Diameter: 1.0 in. (2.54 cm)

Overall Probe Length: 2.5 in. (6.35 cm)

Overmolded Joint Dimensions:

Width: 0.44 in. (1.12 cm)

Height: 0.58 in. (1.47 cm)

Length: 2.25 in. (5.72 cm

Cable Diameter: 0.245 in. (0.622 cm)

Material
Disk:

Anodized

Aluminum

Cable

Jacket:

Santoprene

Cable/Probe Connection: Santoprene

Weight: 0.2 lbs (90.7 g) with 10.5 ft (3.2 m) cable

The black outer jacket of the cable is Santoprene

®

rubber. This

compound was chosen for its resistance to temperature extremes,
moisture, and UV degradation. However, this jacket will support
combustion in air. It is rated as slow burning when tested
according to U.L. 94 H.B. and will pass FMVSS302. Local fire
codes may preclude its use inside buildings.