110PV Surface Temperature Probe

6. Measurement

Understanding the details in this section is not necessary for general operation
of the 110PV Probe with CSI's dataloggers.

G

Ω

Signal

GREEN

Sense +

Volt Excite

WHITE

Sense -

Signal Reference

RED

BLACK

PURPLE

Signal Reference

BLUE

THERMISTOR

CLEAR (shield)

4.99 k , 0.1%

FIGURE 6-1. 110PV Thermistor Probe schematic

Simple half bridge measurement, ignoring cable resistance

The measured voltage, V, is:

t

EX

R

V

V

+

=

990

,

4

990

,

4

Where V

EX

is the excitation voltage, 4,990 ohms is the resistance of the fixed

resistor and R

t

is the resistance of the thermistor

The resistance of the thermistor is:

=

1

990

,

4

V

V

R

EX

t

The Steinhart-Hart equation is used to calculate temperature from Resistance:

3

))

(ln(

)

ln(

1

T

T

K

R

C

R

B

A

T

+

+

=

Where T

K

is the temperature in Kelvin. The Steinhart- Hart coefficients used

are:

A = 1.129241x10

-3

B = 2.341077x10

-4

C = 8.775468x10

-8

18