Campbell Scientific 092 Barometric Pressure Sensor User Manual
Page 18
Appendix A. Analog Output Configurations
A-2
TABLE A-2. Dip Switches SW2: Analog Output Pressure Range Settings
RANGE SELECTION
SWITCH SETTINGS
mbar
mm Hg
in. Hg
LOWER UPPER LOWER
UPPER
LOWER UPPER SW2-1 SW2-2 SW2-3 SW2-4
600 800
450 600
17.72
23.62
ON
ON
ON
ON
600 900
450 675
17.72
26.58
OFF
ON
ON
ON
600 1000
450 750
17.72
29.53
ON
OFF
ON
ON
600
1
1100
1
450
1
825
1
17.72
1
32.48
1
OFF
1
OFF
1
ON
1
ON
1
700 800
525 600
20.67
23.62
ON
ON
OFF
ON
700
900
525
675
20.67 26.58
OFF ON OFF ON
700 1000
525 750
20.67
29.53
ON
OFF
OFF
ON
700
1100
525
825
20.67 32.48
OFF OFF OFF ON
677.1
1083.6
508.0
812.8
20.00
32.00
ON
ON
ON
OFF
800 900
600 675
23.62
26.58
OFF
ON
ON
OFF
800 1000
600 750
23.62
29.53
ON
OFF
ON
OFF
800 1100
600 825
23.62
32.48
OFF
OFF
ON
OFF
2 2
2
2
2
2
ON
2
ON
2
OFF
2
OFF
2
2 2
2
2
2
2
OFF
2
ON
2
OFF
2
OFF
2
900 1000
675 750
26.58
29.53
ON
OFF
OFF
OFF
900
1100
675
825
26.58 32.48
OFF OFF OFF OFF
1
Factory configuration at Campbell Scientific
2
Settings with no pressure values are invalid settings. If the switches are set to either of these combinations,
the analog output will default to the full range of 600 to 1100 mbar.