Specifications, Installation, 1 mounting – Campbell Scientific 092 Barometric Pressure Sensor User Manual
Page 13: Mounting
Model 092 Barometric Pressure Sensor
6. Specifications
Features:
• The following 092 barometer settings are configurable:
o measurement range
o data output option / communication option
o analog output range
• Specifications listed here are limited to those applicable to the 092 as
it is configured and shipped by Campbell Scientific, Inc.
Compatibility: CR800
series
CR1000
CR3000
CR23X
CR10(X)
Measurement
Range:
600 to 1100 mbar
Resolution:
0.1 mbar (.003 Hg)
Accuracy:
±0.35 mbar at 25°C
±0.75 mbar at 0° to55°C
±1.5 mbar at -40°C
Long-term stability:
±1 mbar in 12 months
Temperature
Operating range: -40°
to
55°C
Compensated range: -40°
to
55°C
Power requirement
Source:
6 to 16 Vdc
Load:
10 mA at 12 Vdc
Output:
Analog, 0 to 5 Vdc
Enclosure: Polycarbonate
Size:
120 x 80 x 55 mm (4.7 x 3.1 x 2.2 in)
CE Certification:
Calibration:
Factory-calibrated against a NIST
traceable anemometer.
7. Installation
7.1 Mounting
The 092 is factory-mounted to a solar shield. The solar shield mounts by way
of two u-bolts to a vertical pipe with diameter of up to 2-inch IPS. Install the
sensor such that it faces north in the northern hemisphere, or south in the
7