Appendix A. Analog Output
Configurations

The 092 ships from Campbell Scientific pre-wired and pre-configured for
analog output with the following settings:

Analog Output Voltage: 0 to 5 Vdc

Range: 600 to 1100 mbar

Output voltage and measurement ranges are set by switches. Dip switch
settings are listed in TABLE A-1 and TABLE A-2. Highlighted settings are
those set at the factory by Campbell Scientific. Complete information
concerning measurement, output, and communications settings is available in
the Met One 092, 594, 6633A Barometric Pressure Sensor Manual, which is
available at

www.campbellsci.com

.

TABLE A-1. Dip Switches SW1:

Analog Output Voltage Settings

Range SW1-1 SW1-2

0-1 Vdc

On

On

0-2 Vdc

On

Off

0-2.5 Vdc

Off

On

0-5 Vdc*

Off*

Off*

*Factory configuration at Campbell Scientific.

A-1