beautypg.com

2 maintenance, 3 calibration, Attributions and references – Campbell Scientific 107-L Temperature Probe User Manual

Page 20: Maintenance, Calibration

background image

Model 107 Temperature Probe

Symptom: Temperature is reported as

–86 (°C) or –53 (°F).

Verify the black wire is connected to the switched excitation channel as

specified by the measurement instruction.

Symptom: Incorrect temperature is reported.

Verify the multiplier and offset arguments in the measurement instructions

are correct for the desired units (Section 7.2, Datalogger Programming).

Check the cable for signs of damage and possible moisture intrusion.

Symptom: Unstable temperature is reported.

Most likely a result of electromagnetic interference. Try using the 60 or 50

Hz integration options, and/or increasing the settling time as described in

Section 8.3, Electrically Noisy Environments, and Section 8.4, Long Cable

Lengths. Make sure the clear shield wire is connected to datalogger

ground, and the datalogger is properly grounded.

9.2 Maintenance

The 107 probe requires minimal maintenance. For air temperature

measurements, check the radiation shield monthly to make sure it is clean and

free from debris. Periodically check cabling for signs of damage and possible

moisture intrusion.

9.3 Calibration

Calibration of the 107 probe is not necessary unless the application requires

removal of the thermistor interchangeability offset described in Section 6,

Specifications. If performing the one point calibration with an Edlog

datalogger, be aware of this precaution:

The value of the offset must be chosen so that the probe outputs the

temperature calculated by the polynomial, not the actual calibration

temperature. For example, a 107 probe placed in a calibration chamber at 0 °C

outputs 0.1 °C. An Offset argument of –0.16 is required for Edlog dataloggers

because at 0 °C, the polynomial calculates a temperature of –0.06 °C

(Appendix C, Conversion of Thermistor Resistance or Voltage Ratio to

Temperature).

10. Attributions and References

Santoprene

®

is a registered trademark of Exxon Mobile Corporation.

EPA installation standard: Quality Assurance Handbook for Air Pollution

Measurement Systems – Volume IV: Meteorological Measurements Version 2.0

WMO standard: WMO No. 8, Seventh edition, 6 Aug 2008 Guide to

Meteorological Instruments and Methods of Observation

14