For CRBasic dataloggers, the 60 Hz and 50 Hz integration options include a 3

ms settling time; longer settling times can be entered into the Settling Time

parameter. The following example uses a 20000 µs settling time.

Therm107

(T107_C,1,1,1,20000,_60Hz,1.0,0.0)

Edlog

The 60 Hz and 50 Hz rejection options for the CR10X, CR510, and CR23X

include a delay to accommodate long cable lengths.

When using the CR10, 21X, and CR7 dataloggers, if the 107 probe has cable

lengths of more than 300 feet, use the

Excite-Delay (SE) (P4) instruction with

a 20 ms delay to measure temperature, as shown in the following Edlog code.

The CR10, 21X, and CR7 dataloggers have no programming provision for

compensating for long cables and filtering 60 Hz or 50 Hz noise

simultaneously.

1: Excite-Delay (SE) (P4)
1:

1

Reps

2:

2

7.5 mV Slow Range

3:

9

SE Channel

4:

3

Excite all reps w/Exchan 3

5:

2

Delay (units 0.01 sec)

6:

2000

mV Excitation ;21X,CR7: 4000 mV

7:

1

Loc [ Air_Temp ]

8:

.4

Mult ;21X,CR7: 0.2

9:

0

Offset


2: Polynomial (P55)
1:

1

Reps

2:

1

X Loc [ Air_Temp ]

3:

1

F(X) Loc [ Air_Temp ]

4:

-53.46

C0

5:

90.807

C1

6:

-83.257

C2

7:

52.283

C3

8:

-16.723

C4

9:

2.211

C5

9. Troubleshooting and Maintenance

All factory repairs and recalibrations require a returned material

authorization (RMA) and completion of the “Declaration of

Hazardous Material and Decontamination” form. Refer to the

Assistance

page at the beginning of this manual for more

information.

9.1 Troubleshooting

Symptom: Temperature is reported as

NAN, -INF, or -9999.

Verify the red wire is connected to the correct single-ended analog input

channel as specified by the measurement instruction, and the purple wire is

connected to datalogger ground.