beautypg.com

Specifications, Installation, 1 siting – Campbell Scientific 083E Relative Humidity and Temperature Sensor User Manual

Page 13

background image

Model 083E Relative Humidity and Temperature Sensor

6. Specifications

Features:

• Relative humidity sensing element: thin film polymer capacitor

• Temperature sensing element: thermistor

Compatibility
Dataloggers:
CR800

series

CR1000

CR3000

Measurement

Ranges

RH:

0 to 100% relative humidity

Temperature: -50° to 50°C (-58° to 122°F)

Accuracies

RH:

±2.0% from 0 to 100% RH

Temperature: ±0.10° C (0.18° F)

Response

RH:

10 s with 2 m/s wind aspiration

Temperature

Operating:

-50° to 50°C (-58° to 122°F)

See

Section

8.2,

Limitations of RH Measurement at

Below Freezing Temperatures.

Compensation:

RH is temperature compensated internally

Power requirement
Source:

10 to 18 Vdc

Load:

< 5 mA

Outputs
RH:

0 to 1 Vdc

Temperature: Resistive

bridge

Dimensions
Length:

216 mm (8.5 in)

Diameter:

19 mm (0.75 in)

7. Installation

If the sensor is to be mounted in a radiation shield, refer to the radiation shield
manual section for mounting details. A typical installation is illustrated in the
Section 7.2, Mounting.

Sensors not installed in a radiation shield should be mounted in a representative
location having good airflow and shaded from sunlight or other radiant heat
source.

7.1 Siting

Locate sensors over an open level area at least 9 m (EPA) in diameter. The
surface should be covered by short grass, or where grass does not grow, the
natural earth surface. Locate sensors away from objects at least a distance
equal to four times the height of the objects, and at least 30 m (EPA) from

7