Model ixldp transducer instruction sheet, Figure 1. general dimensions – Ashcroft IXLdp - Low Differential Pressure Transmitter User Manual
Page 2

MOUNTING
FOOT
➤
➤
➤
➤
➤
➤
➤
ø 0.2
1.4
4.31
➤
➤
➤
➤
➤
0.2 TYP
2) PLACES
MOUNTING FOOT
2.68
ACCESS COVER (TERMINAL BLOCK)
➤
➤
MOUNTING FOOT
SPAN POT
ZERO POT
LOCATION
➤
➤
➤
➤
➤
➤
2.7
4.6
➤
➤
0.7
➤
➤
5
➤
1/2 NPSM THREAD TYP BOTH
SIDES FOR CONDUIT PIPING
ACCESS COVER
(ZERO AND SPAN POTS)
3
➤
➤
➤
➤
2.5
➤
➤
➤
➤
➤
➤
➤
➤
0.8
1.34
3.0
0.7
HIGH
LOW
1/4 NPT
PRESSURE
FITTINGS
➤
➤
Figure 1.
General Dimensions
(INCHES)
MODEL IXLdp TRANSDUCER
INSTRUCTION SHEET
OPERATING
REGION
LOOP SUPPLY VOLTAGE (VDC)
V
0
10 12
20
30
36
40
1250
1045
1000
750
500
250
0
LOOP
RESIST
ANCE (OHMS)
LOAD LIMITATION 4-20mA OUTPUT ONLY
F
Figure 2. Loop Supply Voltage vs. Loop Resistance
Figure 3.
Current (4-20mA) Output Wiring
EARTH
GROUND
POWER
SUPPLY
V+
V–
F
R
S
(LOAD)
EARTH
GROUND
SIGNAL
OUT
+
–
V+
V-
F
POWER
SUPPLY
V
min
= 12 + (.022A x R
L
)
R
L
= RS+ RW
R
L
= Loop Resistance (ohms)
R
S
= Sense Resistance (ohms)
R
W
= Wire Resistance (ohms)
*INCLUDES A 10% SAFETY FACTOR
Figure 4.
Voltage Output Wiring
© 2011 Ashcroft Inc., 250 East Main St., Stratford, CT 06614-5145, U.S.A. Tel: 203-378-8281, Fax: 203-385-0499, email: [email protected], www.ashcroft.com