beautypg.com

Is displayed, Select the interval with buttons ▲ or ▼. text – Vaisala HMP46 User Manual

Page 60

background image

HMI41 AND PROBES
Operating Manual

M210316EN-A

56

3.2

Setting the measurement duration for automatic data collecting

RH

set min max

hyst hold

Lo bat Hi

Pabs

Pa

g/m

T

Td

°C°F

°C°F

%

Turn the indicator on. Within 1 - 2 seconds press
the button HOLD until the text

REC AUTO

appears,

then release the button.

Press ENTER when the text

REC AUTO

is displayed.

max

Previously set duration time appears. If the previ-
ously set duration is too long for the current battery
charge, the calculated longest duration time possible
for the current battery charge appears instead; this is
also indicated with the text

MAX

.

bat

Set the duration time with buttons ▲ or ▼. The
duration time can be set from 15 minutes to 7 days.
Text

BAT

on the display indicates that the battery

charge is not sufficient for the duration time
selected; select a shorter duration time.
Turn the indicator off or press ENTER for setting
the measurement interval.

3.3

Setting the measurement interval for automatic data collecting

min

Previously set measurement interval appears. If
the previously set interval is too long for the
indicator's current memory capacity, the calcu-
lated shortest possible interval appears instead;
this is also indicated with the text

MIN

.

Select the interval with buttons ▲ or ▼.
Text

LO

on the display indicates that there is not

enough memory left for the chosen interval;
select a longer interval.
Press ENTER to activate the measurements in
the automatic data collecting mode.

RH

set

T

°C

%

Press ON/OFF if you wish to stop the automatic
data collecting.