HMI41 AND DATA COLLECTING

The HMI41 indicator can also be used for collecting the measurement data.
The data is stored in the indicator's non-volatile memory which means that it
is not lost when the indicator is turned off. Please, note also that the automatic
power-off function is not active during data collecting even if previously
selected (see Chapter 8.3). When the data collecting has ended, the automatic
power-off function becomes active again.

6.1

Entering the data collecting mode

Enter the data collecting mode by turning the indicator on with the ON/OFF
button. The following appears for a couple of seconds:

RH

set min max

hyst hold

Lo bat Hi

P

Pa

h

g/m

T

Td

°C °F

°C °F

%

Release the ON/OFF button and press immediately the button HOLD. The
software version and the probe type indication appear, after which the display
changes automatically to show the battery charge:

bat Hi

Within a couple of seconds the text

REC AUTO

appears on the display; release

the HOLD button.

This is the main display of the data collecting mode. With the button MODE
you can enter the

REC CATCH

mode (manual data collecting, see Chapter 6.2)

and by pressing MODE again, the

REC READ

mode (reading the measurement

results, see Chapter 6.5). By pressing ENTER, you can set the measurement
duration and by pressing ENTER again, the measurement interval (Chapters
6.3 and 6.4). By pressing HOLD, you can always return to the previous
display.