2 current output scale selection mode (i) – Vaisala Calibration of Series HMDW6070 and HMP140 Transmitter with HMI41 User Manual
Page 24

HMI41 Calibration Option
19116ZZ-U218-1.1
Appendix 1
1996-08-26
3
2.2
Current output scale selection mode (I)
DISPLAY
WHAT TO DO
PRESS:
Select the current output scale selection
mode (I)
▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection
set
Select the output
(0...20 mA or 4...20 mA)
▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection
set
Lo
T
T
°C
°C
Set the lower limit of the temperature
range
▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection
set
Hi
T
T
°C
°C
Set the upper limit of the temperature
range
▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection
ON/OFF
See also other documents in the category Vaisala Humidifiers:
- Calibration of Digital Transmitters with HMI41 (36 pages)
- Calibration of Series HMDW2030 and HMP130 Transmitter with HMI41 (14 pages)
- HM34 (30 pages)
- HM40 (47 pages)
- HM44 (52 pages)
- HM70 (83 pages)
- HMD40 (1 page)
- HMD60 (4 pages)
- HMD70 (18 pages)
- HMDW110 (62 pages)
- HMDW80 (51 pages)
- HMI41 (74 pages)
- HMP41 (72 pages)
- HMK15 (39 pages)
- HMM100 (71 pages)
- HMM105 (23 pages)
- HMM211 (42 pages)
- HMM212 (36 pages)
- HMM213 (52 pages)
- HMP140 (28 pages)
- HMP155 (84 pages)
- HMP228 (115 pages)
- HMP230 (163 pages)
- HMP240 (130 pages)
- HMP260 (118 pages)
- HMP60 (71 pages)
- HMT100 (52 pages)
- HMT120 (87 pages)
- HMT130 (95 pages)
- HMT140 (76 pages)
- HMT310 (88 pages)
- HMT310 (105 pages)
- HMT330 (209 pages)
- HMT360 (97 pages)
- HMT360 (63 pages)
- HMT360N (110 pages)
- HMW40 (1 page)
- HMW90 (110 pages)
- SHM40 (68 pages)
- RDP100 (14 pages)