beautypg.com

Calibration, 1 getting started – Vaisala Calibration of Series HMDW6070 and HMP140 Transmitter with HMI41 User Manual

Page 17

background image

HMI41 Calibration Option

19116ZZ-U218en-1.1

Operating Manual

1996-08-29

13

3.

CALIBRATION

For a successful calibration, it is essential that the probe of the HMI41 and
that of the transmitter are at the same temperature, and that the reference
probe has been previously calibrated. Always allow enough time for the
readings to stabilize. Note that the stabilization time depends on the ambient
conditions and may vary from 10 minutes to a couple of hours.

3.1

Getting started

After having selected the desired function (Chapter 1.2) and necessary scale
selections (Chapter 2), turn the HMI41 off and connect the calibration cable to
the EXT connector at the bottom of the HMI41 (see Figure 3.1) and to the
appropriate connector in the transmitter (see Figures 3.2 - 3.4).

EN TER

MO DE

HOLD

ON/OFF

EXT

C o n n e c t o r f o r t h e

c a l i b r a t i o n c a b l e

Figure 3.1

Location of the HMI41 calibration connector

T GAIN

RH GAIN

RH OFFSET

T OFFSET

T
TEST

TEST

RH

HMD60U/Y

T GAIN

RH GAIN

RH OFFSET

T OFFSET

TEST CONNECTOR

U

U

U

U

I-module 0-20mA

1

1

TEST CONNECTOR
FOR CURRENT OUTPUTS

FOR VOLTAGE OUTPUTS

HMD70U/Y

Figure 3.2

Calibration connectors and potentiometers in

the HMD60 and HMD70 transmitters