Calibrate dissolved oxygen in mg/L (ppm) mode
Note...

mg/L (ppm) calibration can only be performed after 100% calibration

has been completed.

1. Clean the probe thoroughly with deionized water.

2. Put the probe into a sample where the oxygen content

is known. The solution should have a flow rate of at

least 0.3m/s.

3. Press

RANGE to select the mg/L(ppm) scale (see Fig. A).

4. Press

CAL to enter calibration mode and begin calibration.

The measurement value will flash (see Fig. B).

5. Press ↑ to increase the Measurement Value.

Hold down ↑ to continually increase the value.

Press ↓ to reduce the Measurement Value.

Hold down ↓ to continually reduce the value.

The Measurement Value range is 0.00 ~ 19.99mg/L.

6. Press

SETUP to save the calibration value and exit to

Normal Mode (see Fig. C).

7. Press

EXIT to return to Normal Mode without saving.