0% Calibration
Note...

The measurement value must read below 10% to calibrate the

meter at 0%. All other keys except POWER ON/OFF and HOLD are

inoperable while performing calibration.

1. Clean the probe thoroughly with deionized water.

2. Press

RANGE to select the % saturation scale (see Fig. A).

3. Place the probe into 0 Oxygen Solution.

The solution should have a flow rate of at least 0.3m/s

4. Press

CAL for 2 seconds to enter the calibration mode

(see Fig. B)

5. Once the reading is steady the display will read

“Finished” (see Fig. C).

6. Press

EXIT to return to Normal Mode (see Fig. D).