Set Barometric Pressure Parameters

1. Press

SETUP to enter the Setup Mode.

2. Press ↑ or ↓ until the Bar. Pressure screen is seen (see Fig. A).

3.

SETUP to enter the Select Bar. Unit screen (see Fig. B or Fig. C).

4. Press ↑ or ↓ to scroll through the available barometric

measurement scales.

5. Press

SETUP to save the displayed measurement scale

and enter the Set Bar. Value screen (see Fig. D or Fig. E).

The current Bar. Value number will flash.

6. Press ↑ to increase the Bar. Value.

Hold down ↑ to continually increase the value.

7. Press ↓ to reduce the Bar. Value.

Hold down ↓ to continually reduce the value.
If the Bar. Value Measurement scale is mmHg the

available range is 500~1499.
If the Bar. Value measurement scale is kPa the

available range is 66.6~199.9

8. Press

SETUP to save the changes and return to Normal Mode.

9. Press

EXIT to return to Normal Mode without saving changes.