4-F. Temperature Unit Default Setting
Press the SET (8) button as needed to reach the "temp. unit"
screen.
Use the ▲ (5) and ▼ (6) buttons to switch between °F and °C.
Press the ENTER (4) button.
Press the SET (8) button to advance to the next option, or press the ESC
(3) button to exit.
4-G. Temperature Compensation Factor (ATC) Default Setting
The ATC Factor is adjustable between 0 to 5.0% per °C.
Press the SET (8) button as needed to reach the
"Temp. Comp." screen.
Use the ▲ (5), ▼ (6), and SEND (7) buttons to set the
ATC Factor.
Press the ENTER (4) button to save the settings.
Press the SET (8) button to advance to the next option, or press the ESC
(3) button to exit.
4-H. Conductivity (uS, mS), TDS (PPM) Default Setting
Press the SET (8) button as needed to reach the "CD/TDS"
screen.
Use the ▲ (5) and ▼ (6) buttons to switch between uS, mS and PPM.
Press the ENTER (4) button.
Press the SET (8) button to advance to the next option, or press the ESC
(3) button to exit.
4-I. Escape from Advanced Adjustments
For the above procedures, press the ESC (3) button before pressing the EN-
TER
(4) button to exit advanced adjustments without saving the changes.
5. UPLOADING DATA FROM THE METER
(Auto data logging readings only—manual readings cannot be upload-
ed)
To upload the data to a computer, connect the RS232 cable and run the soft-
ware (see page 11). You can only send one memory block at a time.
First, cancel the hold and record functions if they are activated.
Press the SEND (7) button for at least 2 seconds until the bottom right dis-
play shows "Transmit Mode."
Release the button.
The following screens flash alternately:
1=PPM
0=uS, mS
Temp. Comp.
^, v SEND(→)
1=°F
0=°C