Sper Scientific 850039 Conductivity TDS Meter - Datalogging User Manual
Page 7
- 7 -
4-A. Check Remaining Number of Data Points
Press the SET (8) button at least 2 seconds until
the lower display shows:
"
XXXXX" is the number of free data points.
4-B. Clear Memory
Press the SET (8) button as needed to reach
the "clear memory" screen. The number of saved data points is displayed.
To delete saved data from memory, press the ENTER (4) button once,
then press the ENTER (4) button again to confirm. The display shows:
"0 Ok ESC:Quit".
If you DO NOT want to clear the memory, press the ESC (3) button to exit,
OR press the SET (8) button to advance to the next option.
4-C. Date/Time Setting
Press the SET (8) button as needed to reach
the "date/time set" screen.
Use the ▲ (5), ▼ (6), and ENTER (4) buttons
to select the year, month, date, hours, minutes and seconds.
Press the ENTER (4) button to save the settings.
Press the SET (8) button to advance to the next option, or press the ESC
(3) button to exit.
4-D. Sample Time Setting
Press the SET (8) button as needed to reach
the "sample time" screen.
Use the ▲ (5), ▼ (6), and ENTER (4) buttons
to select the sample time (hour-min-sec).
Press the ENTER (4) button to save the settings.
Press the SET (8) button to advance to the next option, or press the ESC
(3) button to exit.
4-E. Auto Power Off Default Setting
The meter will automatically shut off after about
10 minutes without activity.
Press the SET (8) button as needed to reach
the "auto power off" screen.
Use the ▲ (5) or ▼ (6) buttons to enable / disable this feature.
Press the ENTER (4) button to save the settings.
Press the SET (8) button to advance to the next option, or press the ESC
(3) button to exit.
1=Auto Power Off - Enabled
0=Auto Power Off - Disabled
XXXXX Memory Space
06:01:03 Date/Time Set
y-m-d ^, v ENTER
(→)
00:00:02 Sample Time
h-m-s ^, v ENTER(→)
Clear Memory
Esc:N Enter:Y
107