Page 8

background image

3)Insert the " Testing bottle " to the bottom of

" Container " ( 3-2, Fig. 1 ) completely.

4)Before the measurement, it should envelop in the

" Cover " ( 3-1, Fig. 1 ) completely.

Remark :

Before the measurement, it should

keep the outside of Testing Bottle

under the dry condition and without

existing any dust.

4-2 Measurement

Before the measurement, should select the

measurement model to " Free Chlorine " or " Total

Chlorine " first, the setting procedures refer to

chapter 4-3, page 8.

ZERO setting for the liquid

1)Power ON the meter by pressing the " Power Button "

( 3-6, Fig. 1 ) once.

The Display ( 3-3, Fig. 1 ) will show the text " Free (
TOTAL ) " approx. 1 second. then show " CAL 0 ", now
the meter for the ZERO setting procedures.

2)Fill the measurement liquid into the " Testing bottle "

( 3-11, 3-12, Fig. 1 ) .

Attention :

Model 0601

It should fill the

10 mL

measurement

liquid capacity

until its level

Fig. 3

up to the " level

marker " of the

" Testing bottle "

6