beautypg.com

Calibration procedure – Sper Scientific LUCL-2006 Chlorine Meter User Manual

Page 12

background image

5. CALIBRATION PROCEDURE

1)The meter can be calibrated under following calibration

points :

Free chlorine

*Zero
*1.00 ppm

or

Total chlorine

*Zero
*1.00 ppm

2)The meter ship along with

* Zero Chlorine standard solution ( CL-01 ) X 1 PC,
* 1.0 ppm Free Chlorine standard solution ( CF-01 ) X 1 PC.
* 1.0 ppm Total Chlorine standard solution ( CT-01) X 1 PC.

as the standard accessories.

3)The complete calibration should be executed by

following two solution :

*Zero Chlorine standard solution
*1.0 ppm standard solution ( Free or Total ).

10